EasyManuals Logo

Hioki MEMORY HiCORDER MR8880 User Manual

Hioki MEMORY HiCORDER MR8880
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #243 background imageLoading...
Page #243 background image
13.1 MR8880-20 Specifications
233
Chapter 13 Specifications
13
(8) Trigger Function
Trigger method Digital comparison
Repeat recording Single, Repeat
Trigger timing
• High-speed function: Start
• Real-time function: Start, stop, start&stop (conditions can be set
for each of start and stop)
Trigger criteria AND or OR possible between all trigger sources
Trigger source
Trigger source selectable for each channel
Free-running when all trigger sources are Off
1. Analog inputs: CH1 to CH4
2. Logic inputs: LA1 to LA4, LB1 to LB4 (4 channels x 2 probes)
3. External trigger: Input signal to external trigger terminal
Interval trigger
Fixed interval recording at specified measurement interval (day,
hour, minute, second) is possible.
A trigger is applied simultaneously with start of measurement, and
then a trigger is subsequently applied at the set measurement
interval. (Use the schedule function to specify the start time.)
Trigger types
(analog)
1. Level trigger
A trigger is applied when rise or fall to set voltage value.
2. Window trigger
Set the upper and lower limits of trigger level
3. Voltage drop trigger ([Highspeed] only)
A trigger is applied when the voltage peak falls below the set
level.
(Dedicated for commercial power supply of 50/60 Hz.
The trigger can not be set with the time base range of 50 ms/div
and 100 ms/div.)
4. Waveform judgment trigger ([Highspeed] only)
A trigger for performing waveform judgment in real time.
(Dedicated for commercial power supply. The trigger can not be
set with the time base range of 50 ms/div and 100 ms/div.)
Comparison targets: Sine wave and previous waveform
Judgment period: 20 µs (with 100 μs to 2 ms/div), sampling
period (with 5 ms to 20 ms/div)
Maximum sampling speed: 1 MS/s (with time base of 100 µs/
div)
Trigger types (logic)
Pattern trigger: Settable to 1/0/x for each input (ignore x)
• Configure the pattern setting for each probe.
• The trigger condition (AND/OR) can be set between logic input
channels in each probe.
• The trigger condition between probes is in accordance with over-
all trigger condition (AND/OR).

Table of Contents

Other manuals for Hioki MEMORY HiCORDER MR8880

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiCORDER MR8880 and is the answer not in the manual?

Hioki MEMORY HiCORDER MR8880 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiCORDER MR8880
CategoryVoice Recorder
LanguageEnglish

Related product manuals