4.8 Viewing Motor Measurement Values (Model PW3390-03 only)
104
You can set the phase zero adjustment correction value as desired.
The phase zero adjustment correction value can be set as desired within the range of -180.00
° to
+180.00
°. Enter the phase difference between the rotation input signal pulse and the U1 fundamental
component.
If using the instrument in an environment in which the phase angle is expressed as a value from 0
° to
360
°, enter after converting to a value from -180° to +180°.
Setting the phase zero adjustment correction value manually
Phase zero adjustment is valid only when CH B is set to pulse input and the [Harmonic sync
source] is set to [Ext]. No setting can be entered if CH B is not set to pulse input.
The screen displays the present phase zero adjustment correction value. Consequently,
pressing the key while holding down the key on the Measurement screen to
perform zero adjustment will cause the correction value to be overwritten. In addition, reset-
ting the correction value by pressing the key on the Measurement screen while hold-
ing down the key will revert the correction value to 0.
The set phase zero adjustment correction value will be subtracted from pulse-based voltage
and current phase measured values.
Setting Procedure
Select with the F keys
Select
[ Phase Adjust]
F1
Increase/reduce
value
F2
F3
Select digit to
change
F4
F5
Change sign
Display the
[Motor] page
Press