129
6
Chapter 6 Changing System Settings
Start page Select the screen to appear when the instrument is turned on.
Zero suppress This setting establishes a level below which values are treated as zero for data
acquisition purposes.
Clock Sets the internal real-time clock. Data is saved and managed according to this
setting.
CSV file format Sets the CSV file format. This setting applies to measurement data saved manu-
ally as well as all measurement data and waveform data saved automatically.
Wiring Initially display the wiring screen.
Last scr
Initially displays the Measurement screen that was displayed
when the instrument was turned off.
OFF Zero suppression is disabled. To display small values, disable
zero suppression.
0.1% f.s./0.5% f.s. Measured values below the selected level are treated as zero.
+1/-1 Increments/decrements by 1.
+10/-10 Increments/decrements by 10.
Set Press to apply setting changes (resets seconds to 00).
CSV Use commas (“,”) as delimiters in measurement data.
Use periods (“.”) as decimal points.
SSV Use semicolons (“;”) as delimiters in measurement data.
Use commas (“,”) as decimal points.