7.5 Measurement Data Saving
142
Timing controls available for auto-saving are as follows.
Auto-Saving Operations
Interval Timing Control
Save Save Save Save
Interval Interval Interval
Save
Timer Control
Save Save
Timer Setting
Auto Stop
Real-Time Clock Timing
Control
Save Save
Waiting Time Between Clock Start/Stop Settings
Auto Start
Specified Start Time
Auto Stop
Specified Stop Time
Timer + Interval Timing
Control
Save Save Save SaveSave
Timer Setting
Interval Interval Interval Interval
Auto Stop
Clock + Interval Timing
Control
Save Save SaveSave
Waiting
Interval Interval Interval
Auto Start
Specified Start Time
Auto Stop
Specified Stop Time
Time Between
Clock Start/Stop
Settings
•
Settings cannot be changed while timing controls are enabled. Also, when auto-ranging is
enabled along with timing control, the range that is active when is pressed remains fixed.
• The all data is saved in the same file under timing control.
When integration is reset, data is saved to a new file at the next start time.
• When the timer stop time and the interval ending time do not match, the timer stop time
has priority, and the last interval is truncated.
• When the clock control stop time and the interval ending time do not match, the clock
control stop time has priority, and the last interval is truncated.
• When the storage media becomes full during auto-saving, an error is displayed and sav-
ing stops. In that case, auto-saving can be resumed (using an automatically named file
with the same name) by replacing the CF card with another (formatted) CF card.
To OFF the interval, see (p. 109)