8.1 Connecting Multiple PW3390 (Synchronized Measurements)
159
8
Chapter 8 Connecting External Devices
Set each instrument to be either the master or a slave.
These settings are made on the [Interface] page of the Setting screen.
Sync event Select the events to be synchronized
(master settings are reflected on the slaves)
Instrument Settings for Synchronous Measurement
Setting Procedure
Display the
[Interface] page
Select with the F keys
Select
[Sync event]
Select with the F keys
Reboot the instrument when
changing from master to slave.
Select
[Sync control]
HOLD Pressing on the master activates Data Hold on all instruments.
SAVE Pressing on the master executes manual saving on all instruments.
COPY Pressing + on the master captures the screen image on all
instruments.
The RTC clock time, timer, and clock control start and stop times cannot be set on the
slave instruments.
Selecting [SAVE] or [COPY] as a synchronized event sets the manual save destina-
tion folder appropriately and records data on each PW3390.
See "7.5.1 Manually Saving Measurement Data" (p. 139), "7.7 Saving Screen Capture Images" (p.
147)
To save measurement data to storage media with an interval time control combina-
tion, set the same interval setting on the master and all slaves, and enable auto-sav-
ing (set to ON). In this case, selecting [SAVE] as a synchronous event has no effect.
See "5.1 Timing Control Functions" (p. 107), "7.5.2 Auto-Saving Measurement Data" (p. 141)
Confirm no error display on the slave unit's screen when executing the synchroniza-
tion event.