EasyManua.ls Logo

Hioki PW3390 - Page 176

Hioki PW3390
248 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.3 Using Analog and Waveform D/A Output
168
Output signals are waveforms of the instantaneous values of input voltages and currents.
Voltage inputs and current sensor inputs are mutually isolated.
Combine with an oscilloscope to observe waveforms of phenomena such as device inrush current.
About Waveform Outputs
Specifications
Output voltage Select from either ±1 V or ±2 V, crest Factor 2.5 or higher
Output impedance 100 ±5
Output update rate 500 kHz
D/A1: U1, D/A2: I1, D/A3: U2, D/A4: I2, D/A5: U3, D/A6: I3, D/A7: U4, D/A8: I4
Waveform clipping occurs at approximately
±7 V.
Maximum output voltage that can possibly be output from malfunction, etc. is approximately
±12 V.
When using VT or CT ratio, the instrument outputs a voltage that is obtained by multiplying
the range by the VT or CT ratio.
Waveform output consists of uninterrupted instantaneous values, regardless of data hold,
peak hold, or averaging operations.
When auto-ranging is enabled, the analog output levels change with auto-ranging. Be careful
to avoid range conversion mistakes when measuring rapidly fluctuating values. Such mis-
takes can be avoided by using a fixed range. It is recommended to use a fixed range for this
type of measurement.

Table of Contents

Other manuals for Hioki PW3390

Related product manuals