11.3 Error Indication
225
Chapter 11 Maintenance and Service
11
Contact your authorized Hioki distributor or reseller, if a repair should become necessary.
Different sensors! Cannot
change the wiring in the set-
ting file.
Attempted to load an incompatible setting
configuration file.
Settings cannot be loaded unless instrument
options and saved items are the same as those
installed and selected when saved.
See "7.10 Reloading Setting Configurations"
(p. 150)
D/A output function is differ-
ent.
Attempted to load an incompatible setting
configuration file.
Motor analyzing function is
different.
Attempted to load an incompatible setting
configuration file.
Inconsistent items to save
Attempted to load an incompatible setting
configuration file.
CF card error!
This card is not supprted.
Incompatible CF card found.
Use a Hioki CF card option.
See "Chapter 7 Data Saving and File Opera-
tions" (p. 133)
USB memory stick error!
This memory stick is not sup-
prted.
Incompatible CF card found.
Use a Hioki CF card option.
See "Chapter 7 Data Saving and File Opera-
tions" (p. 133)
Failed to write. Writing to storage media failed.
Try again.
Failed to read. Reading from storage media failed.
Failed to save while calculat-
ing the waveform data
Attempted to save a waveform while it is
being created.
Try again after the waveform is created (when
the timer mark disappears).
Failed to create a file. File creation failed for unknown reason.
Try again.
Failed to create a folder. Folder creation failed for unknown reason.
Synchronized signals cannot
be detected.
Synchronized signals cannot be detected
from the master when setting the slave.
Confirm that the master is connected with syn-
chronized cable and that the master is turned
ON.
See "8.1 Connecting Multiple PW3390 (Syn-
chronized Measurements)" (p. 157)
When not using the synchronization function,
set the synchronization control setting to [Mas-
ter].
Unknown error! An unknown error has occurred.
Clear this error by pressing any key except
or once. If the error recurs, contact your
authorized Hioki distributor or reseller.
Error display Cause Remedy
If the measurement lines are energized when the instrument is turned on, the instru-
ment may be damaged or an error message may appear, so before energizing the
lines, turn the instrument on and confirm that no error message is displayed.