9.7 Message Reference
123
9
Messages specific to the RS-232C or GP-IB interface are identified by their corresponding symbols
(/).
(1) System Data Command
(2) Internal Operation Command
Standard Commands
Queries device ID
Syntax
Query
IDN?
Response
<Manufacturer's name>,<Model name>,0,<Software version>
Example
HIOKI,RM3542,0,V1.00
The Device ID is HIOKI RM3542, 0, software version 1.00.
For model RM3542-01, the <Model name> is RM3542-01.
Note
The response message has no header.
Initialize Device
Syntax
Command
RST
Description
Command Resets the instrument to its initial state.
Note
The communications state is not initialized.
Initialized settings are not backed up.
The following methods are available if backup is needed:
Method 1. Use the :SYSTem:BACKup command
Method 2. Use the
:SYSTem:RESet command
Execute Self-Test and Query the Result
Syntax
Query
TST?
Response <0 to 7 (NR1)>
Description
Perform instrument self-test and return the result as numerical value 0 to 7.
Returns zero when no error occurs.
Example
TST?
4
A non-volatile memory error occurred.
Correct measurement may not be possible. Obtain repair before further use.
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
unused unused unused unused unused
non-volatile
memory
RAM ROM