Appendix 8 Table of Commands Compatible with the ADEX AX-162D
A14
Appendix 8 Table of Commands Compatible
with the ADEX AX-162D
Description
Model AX-162D (Original mode) Model RM3542
Command Query Data Command Query Data
Measurement
mode setting
Mm
m = 0/ D (%
mode)/ 1/ R (R
mode)
:CALCulate:LIMit:MODE
m
m = REF%
(REF mode)/
ABS (ABS
mode)
M? XANS1,Mm
:CALCulate:LIM-
it:MODE?
m
Measurement
range setting
RNGr
r = 0 to 9
:RESistance:RANGe r
r = 0 to 120E+6
RNG?
XANS1,RNG
r
:RESistance:RANGe?
r
Reference
resistance
setting
RSfloat
float = 500E-5 to
10900E+4
:CALCulate:LIMit:REF-
erence
float
float = 0 to
120.000E+6
RS?
XANS1,RS
float
:CALCulate:LIMit:REF-
erence?
float
Setting %
mode +/-
tolerances
LHfloat1
float1 = -9999E-
2 to 9999E-2
:CALCulate:LIMit:PER-
Cent
float1,float2
:CALCulate:LIMit:PER-
Cent?
float1,
float2
float1 = -99.990
to 99.990
float2 = -99.990
to 99.990
LH?
XANS1,LH
float1
LLfloat2
float2 = -9999E-
2 to 9999E-2
LL?
XANS1,LL
float2
Setting ABS
mode
thresholds
LHfloat1
float1 = 00001E-
5 to 20000E+4
:CALCulate:LIMit:ABS
float1,float2
:CALCulate:LIMit:ABS?
float1,
float2
float1 = 0 to
120.000E+6
float2 = 0 to
120.000E+6
LH?
XANS1,LH
float1
LLfloat2
float2 = 00001E-
5 to 20000E+4
LL?
XANS1,LL
float2
Increment
(shift range
up) command
SFTRNGU
(Multiple commands)
r1 = r x 10
:RESistance:RANGe?
r
:RESistance:RANGe r1
Decrement
(shift range
down)
command
SFTRNGD
(Multiple commands)
r1 = r/10
:RESistance:RANGe?
r
:RESistance:RANGe r1
Measurement
speed setting
Ss
s = 0 (SLOW
mode)/ 1 (FAST
mode)
:SPEEd s
s = FAST/ ME-
Dium/ SLOW
S?
XANS1,S
s
:SPEEd?
s
Trigger
E TRG
Note: The RM3542 cannot accept continuous commands (a delimiter is necessary between multiple commands).