9.8 Data Exporting Methods
148
Basic Data Exporting Methods
Flexible data exporting is available depending on the application.
9.8 Data Exporting Methods
Free-Run Data Exporting
Initial Setup
:INITiate:CONTinuous ON (continuous measurement enable)
:TRIGger:SOURce IMMediate (internal triggering)
Exporting
:FETCh?
Imports the most recent measurement.
Exporting by Host Triggering
Initial Setup
:INITiate:CONTinuous OFF (continuous measurement disable)
:TRIGger:SOURce IMMediate (internal triggering)
Exporting
:READ?
A trigger occurs, and a measurement is taken and the result is transferred.
Exporting by Pressing the F4 [MANU] Key
*1
or Applying TRIG Signal
Initial Setup
:INITiate:CONTinuous OFF (disable continuous measurement)
:TRIGger:SOURce EXT (external triggering)
Exporting
:READ?
When triggered by the F4 [MANU] Key
*1
or TRIG signal, a measurement is taken and the
result is transferred.
*1. The F4 [MANU] key indicator is displayed when [TRG: EXT] is selected on the Basic Setting screen.