9.7 Message Reference
122
Messages specific to the RS-232C or GP-IB interface are identified by their corresponding symbols
(/).
9.7 Message Reference
Message Reference Interpretation
Shows the command description.
Shows the message syntax.
Explains the command data or
response message.
Describes the message.
Shows an example of an actual
command application.
(Normally described with HEADER
ON,
(except the HEADER command
itself)).
Read/Write the Standard Event Status Enable Register (SESER)
Syntax
Command
Query
Response
ESE <0 to 255 (NR1)>
ESE?
<0 to 255 (NR1)>
Description
Command
The SESER mask is set to the numerical value 0 to 255.
The initial value (at power-on) is 0.
Query
The contents of the SESER, as set by the ESE command,
are returned as an NR1 value (0 to 255).
Example ESE 36
(Sets bits 5 and 2 of SESER)
128 64 32 16 8 4 2 1
bit 7 bit 6 bit 5 bit 4 bit 3 bit 2 bit 1 bit 0
PON URQ CME EXE DDE QYE RQC OPC
< >: Indicates the contents (character or numeric parameters)
of the data portion of a message.
Character parameters are returned as all capital letters.
Numeric Parameters:
NRf Number format may be any of NR1, NR2 and NR3
NR1 Integer data(e.g.: +12, -23, 34)
NR2 Fixed-point data(e.g.: +1.23, -23.45, 3.456)
NR3 Floating-point exponential representation data
(e.g.: +1.0E-2, -2.3E+4)
Command, Query
Response