11.2 Cleaning
180
Before Returning for Repair
If abnormal operation occurs, check the following items.
Symptom Check Items
The display does not ap-
pear when you turn the
power on.
Is the power cord disconnected?
Are connections made correctly?
Verify that the power cord is con-
nected properly.
(p. 22)
Keys do not work. Is any key being held down?
Is Key-Lock enabled (M.LOCK or F.LOCK dis-
played)?
Is remote control communication enabled (RMT dis-
played)?
Is the EXT I/O KEY_LOCK signal asserted (Low
level)?
Verify key operation.
Cancel Key-Lock (p. 60)
Switch to local control (p. 104)
De-assert the KEY_LOCK
signal
(High level).
Measured values are
unstable.
Is the measurement object a power transformer or
other large inductance?
Are the cables and measurement object adequately
shielded?
"Appendix 3 Unstable Measure-
ment Values" (p. A3)
Measured values are
shifted.
Is zero-adjust enabled?
Are the CUR and POT terminals connected before
contacting the measurement object?
Is the measured object calibrated correctly?
Is the GUARD terminal drawing more than 10 mA?
(the GUARD terminal shield may be contacting a
BNC terminal)
There may be a large thermal emf.
"3.6 Zero Adjustment" (p. 32)
If the cause is unknown Try performing a system reset (p. 67). All settings are returned to their factory defaults.)
11.2 Cleaning
To clean the instrument, wipe it gently with a soft cloth moistened with water or mild deter-
gent. Never use solvents such as benzene, alcohol, acetone, ether, ketones, thinners or
gasoline, as they can deform and discolor the case.
Wipe the LCD gently with a soft, dry cloth.