9.7 Message Reference
147
9
Set and Query Contact Check Threshold
Syntax
Command
:RESistance:CONTactcheck:LEVel <Range>,<Level>
Query
Response
:RESistance:CONTactcheck:LEVel? <Range>
<Level>
Command
:LPResistance:CONTactcheck:LEVel <Range>,<Level>
Query
Response
:LPResistance:CONTactcheck:LEVel? <Range>
<Level>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Level> = L1/ L2/ L3 / L4/ L5/ L6/ L7
Contact Check Error Threshold:
(L1: 50
/ L2: 100/ L3: 150/ L4: 200/ L5: 300/ L6: 400/
L7: 500
)
Description
Sets the error threshold value for contact checking.
Set and Query Voltage Level Monitor Level
Syntax
Command
:RESistance:VMONitor <Range>,<1/ 0/ ON/ OFF>
Query
Response
:RESistance:VMONitor? <Range>
<ON/ OFF>
Command
:LPResistance:VMONitor <Range>,<1/ 0/ ON/ OFF>
Query
Response
:LPResistance:VMONitor? <Range>
<ON/ OFF>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
Description
Enables/disables the voltage level monitor function.
Set and Query Voltage Level Monitor Level
Syntax
Command
:RESistance:VMONitor:LEVel <Range>,<Level>
Query
Response
:RESistance:VMONitor:LEVel? <Range>
<Level>
Command
:LPResistance:VMONitor:LEVel <Range>,<Level>
Query
Response
:LPResistance:VMONitor:LEVel? <Range>
<Level>
<Range> = (:RESistance) RNG100MIL/ RNG1000MIL/ RNG10/ RNG100/ RNG1000/ RNG10K/
RNG100K/ RNG1000K/ RNG10MEG/ RNG100MEG
(:LPResistance) RNG1000MIL/ RNG10/ RNG100/ RNG1000
<Level> = L1/ L2/ L3
(L1: Loose/ L2: Normal/ L3: Severe)
Description
Sets the error level for the voltage level monitor function.