Chapter 8 ADJUSTMENT MODE
4.
Press the [SPAN] button when displaying values are stabilized..
Then, check that the A/D display field shows “95000”, and the SPAN A/D display field
shows ”75000”.
5.
If the A/D and SPAN A/D values are not “95000” and ”75000” respectively, remove the
weights from the platter.
Then, place them on the platter again and press the [SPAN] button to check that the A/D
display field shows “95000”, and the SPAN A/D display field shows ”75000”.
UNI-7 Service Manual 8-17