Model 6485 Picoammeter Instruction Manual SCPI Signal Oriented Measurement Commands 12-3
Programming example The following command sequence selects and congures
Model 6485 for one-shot measurements. Each subsequent READ? triggers a single
measurement and requests the reading.
CONF:CURR ' Perform CONFigure operations.
READ? ' Trigger measurement and request reading.
B) FETCh? Request latest reading
This command requests the latest post-processed readings. After sending this command
and addressing Model 6485 to talk, the readings are sent to the computer. This command
does not affect the instrument setup.
This command does not trigger a measurement. The command simply requests the last
group of readings. Note that this command can repeatedly return the same readings. Until
there is a new reading(s), this command continues to return the old reading(s). If your
application requires a fresh reading, use the :READ? command.
This command is automatically asserted when the :READ? or :MEASure? command is
sent.
C) READ? Trigger measurement(s) and request reading(s)
This command is used to trigger and acquire readings. The number of readings depends on
how the trigger model is congured. For example, if congured for 20 measurements (arm
count 1, trigger count 20), 20 sets of readings will be acquired.
When this command is sent, the following commands execute in the order they are pre-
sented:
INITiate
FETCh?
If the instrument is in the idle state, INITiate takes the instrument out of idle to perform
the programmed number measurements. If the instrument is not in the idle state, execution
of this command will wait until it goes back into idle.
The FETCh? command is executed to acquire the reading(s). The readings are sent to the
computer when Model 6485 is addressed to talk.
NOTE If the instrument is programmed to perform an infinite number of measurements
(arm count or trigger count set to infinite), you cannot use the READ? command
to trigger and acquire readings. Use INITiate to start (trigger) the measurement
process, send ABORt to abort the measurement process, and then use
SENSe:DATA[:LATest]? to return the last (latest) reading.

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)