15-2 Performance Verification Model 6485 Instruction Manual
Introduction
Use the procedures in this section to verify that Model 6485 accuracy is within the limits
stated in the instruments one-year accuracy specications. You can perform these verica-
tion procedures:
When you rst receive the instrument to make sure that it was not damaged during
shipment.
To verify that the unit meets factory specications.
To determine if calibration is required.
Follow calibration to make sure it was performed properly.
WARNING The information in this section is intended only for qualified service
personnel. Do not attempt these procedures unless you are qualified to
do so. Some of these procedures may expose you to hazardous voltages,
which could cause personal injury or death if contacted. Use standard
safety precautions when working with hazardous voltages.
NOTE If the instrument is still under warranty, and its performance is outside specified
limits, contact your Keithley representative or the factory to determine the cor-
rect course of action.
Verication test requirements
Be sure that you perform the verication tests:
Under the proper environmental conditions.
After the specied warm-up period.
Using the correct line voltage.
Using the proper test equipment.
Using the specied test signals and reading limits.
Environmental conditions
Conduct your performance verication procedures in a test environment with:
An ambient temperature of 65˚ to 82˚F (18˚ to 28˚C).
A relative humidity of less than 70% unless otherwise noted.