1-16 Getting Started Model 6485 Picoammeter Instruction Manual
Remote setup operation
Saving and restoring user setups
The *SAV and *RCL commands are used to save and recall user setups. These commands
are documented in Section 9.
Restoring factory or GPIB default setups
The SYSTem:PRESet command returns Model 6485 to the factory defaults and the *RST
command returns it to the GPIB defaults. The *RST command is documented in Section
11 and SYSTem:PRESet is covered in Section 12.
Selecting power-on setup
The SYSTem:POSetup command is used to select which setup to return to on power-up.
The SYSTem:POSetup command is documented in Section 12.
Table 1-3
Default settings
Setting
Factory
(:SYStem:PRESet)
GPIB
(*RST)
Trig Layer (CONF-TRIG):
TRIG:
Arm-In Source Event
TRIG-IN
IMM
*
*
Arm Layer (CONF-ARM):
Arm-In Source Event
Arm Count
Input Trigger Link Line
Source Bypass
Output Trigger Link Line
Output Trigger
IMM
INF
1
NEVER
2
Off
*
1
*
*
*
*
Buffer (STORE):
Count
Disabled
No effect
*
*
Digital Filter (AVG):
Count
Type
Advanced:
Noise Tolerance
Off
10
Moving
No (disabled)
0.0%
*
*
*
*
*
Display Resolution (DIGITS) 5½-digits *
Format byte order Swapped Normal
*This factory (:SYStem:PRESet) and bus (*RST) GPIB defaults are the same. Bus settings that are different
from factory reset are as shown.