Model 6485 Picoammeter Instruction Manual IEEE-488 Bus Overview F-9
Uniline commands
ATN, IFC, and REN are asserted only by the controller. SRQ is asserted by an external
device. EOI may be asserted either by the controller or other devices depending on the
direction of data transfer. The following is a description of each command. Each command
is sent by setting the corresponding bus line true.
REN (Remote Enable) REN is sent to set up instruments on the bus for remote opera-
tion. When REN is true, devices will be removed from the local mode. Depending on
device conguration, all front panel controls, except the LOCAL button (if the device is so
equipped), may be locked out when REN is true. Generally, REN should be sent before
attempting to program instruments over the bus.
EOI (End or Identify) EOI is used to positively identify the last byte in a multi-byte
transfer sequence, thus allowing data words of various lengths to be transmitted easily.
IFC (Interface Clear) IFC is used to clear the interface and return all devices to the
talker and listener idle states.
ATN (Attention) The controller sends ATN while transmitting addresses or multiline
commands.
SRQ (Service Request) SRQ is asserted by a device when it requires service from a
controller.
Universal multiline commands
Universal commands are those multiline commands that require no addressing. All devices
equipped to implement such commands will do so simultaneously when the commands
are transmitted. As with all multiline commands, these commands are transmitted with
ATN true.
LLO (Local Lockout) LLO is sent to the instrument to lock out the LOCAL key and all
their front panel controls.
DCL (Device Clear) DCL is used to return instruments to some default state. Instru-
ments usually return to their power-up conditions.
SPE (Serial Poll Enable) SPE is the rst step in the serial polling sequence which is
used to determine which device has requested service.
SPD (Serial Poll Disable) SPD is used by the controller to remove all devices on the
bus from the serial poll mode and is generally the last command in the serial polling
sequence.

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)