16-6 Calibration Model 6485 Instruction Manual
Aborting calibration
You can abort the calibration procedure at any time by pressing the EXIT key.
Current calculations
When calibrating the 2nA-2µA current ranges, you must calculate the actual current val-
ues from the applied calibrator voltages and the characterized Model 5156 Calibration
Standard resistor values. Calibration currents are calculated as follows:
I = V/R
Where: I = required calibration current
V = calibrator voltage
R = actual standard resistor value
For example, assume you are calibrating the 20nA range using a 2V calibrator voltage
with an actual 100.5M standard resistor value. The actual calibration current is:
2V/100.5M = 19.9005nA.
Calibration procedure
The calibration procedure should be performed in the following order:
Preparing for calibration
Offset voltage calibration
Current calibration
Entering calibration dates and saving calibration
Locking out calibration
Preparing for calibration
1. Turn on the Model 6485 and the calibrator, and allow them to warm up for at least
one hour before performing calibration.
2. Press MENU, select CAL, then press ENTER. The instrument will display the fol-
lowing:
CAL: VOFFSET
3. Use the up or down RANGE key to display the following:
CAL: UNLOCK
4. Press ENTER. The instrument will prompt for the calibration code:
CODE? 000000

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)

Related product manuals