I-6 Applications Guide Model 6485 Picoammeter Instruction Manual
Figure I-3
Simplified model of a feedback picoammeter
Source capacitance
DUT source capacitance will also affect the noise performance of the Model 6485 picoam-
meter. In general, as source capacitance increases, the noise also increases. To see how
changes in source capacitance can affect noise gain, again refer to the simplied picoam-
meter model in Figure I-3. The elements of interest for this discussion are the source
capacitance, C
S
and the feedback capacitance C
F
. Taking into account the capacitive reac-
tance of these two elements, the previous noise gain formula must be modied as follows:
Table I-1
Minimum recommended source resistance values
Range Minimum Recommended Source Resistance
2nA, 20nA 1 M to 1 G
200nA, 2µA 10 k to 10 M
20µA, 200µA 100 to 100 k
2mA, 20mA 10 to 1k
R
F
C
F
-
+
V
noise
R
S
C
S
V
S
V
O
Z
F
Z
S
6485
Picoammeter
Current
Source
Output V
NOISE
Input V
NOISE
1Z
F
/Z
S
+()=

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)