Model 6485 Picoammeter Instruction Manual Applications Guide I-17
Figure I-13
Zero check transient
Table I-2
Internal impedance for zero check transient
Range Zcheck
Transient
impedance
2mA,20mA 500Ω
20µA,200µA 50kΩ
200nA,2µA 3.5MΩ
2nA,20nA 11MΩ