Model 6485 Picoammeter Instruction Manual Range, Units, Digits, Rate, and Filters 4-7
Programming example rate
The following command sets the integration rate for all measurement ranges to 2 PLC:
CURR:NPLC 2 ' Set integration rate to 2 PLC.
Filters
Filtering stabilizes noisy measurements caused by noisy input signals. The Model 6485
uses two types of lters: median and digital. The displayed, stored or transmitted reading
is simply the result of the ltering processes. Note that both the median and digital lters
can be in effect at the same time.
With both lters enabled, the median lter operation is performed rst. After the median
lter yields a reading, it is sent to the stack of the digital lter. Therefore, a ltered reading
will not be displayed until both lter operations are completed.
The settings for the lter are global. The MEDN key is used to control the median lter.
The AVG key is used to control the digital lter (press CONFIG and then the AVG key to
congure). When either the median or digital lter is enabled, the FILT annunciator is on.
Median lter
The median lter is used to determine the "middle-most" reading from a group of readings
that are arranged according to size. For example, assume the following readings:
20mA, 1mA, 3mA
The readings are rearranged in an ascending order as follows:
1mA, 3mA, 20mA
From the above readings, it is apparent that 3mA is the median (middle-most) reading.
The number of sample readings used for the median calculation is determined by the
selected rank (1 to 5) as follows:
Sample readings = (2 × R) + 1
where; R is the selected rank (1 to 5)
For example, a rank of 5 will use the last 11 readings to determine the median; (2 × 5) + 1
= 11. Each new reading replaces the oldest reading, and the median is then determined
from the updated sample of readings.
Median filter operation The median lter operates as a moving type lter. For exam-
ple, if the median lter is congured to sample 11 readings (Rank 5), the rst ltered read-
ing will be calculated (and displayed) after 11 readings are acquired and placed in its lter
stack. Each subsequent reading will then be added to the stack (oldest reading discarded)
and another median lter reading will be calculated and displayed. The median lter oper-
ation will reset (start over) whenever the Zero Check operation is performed or the range
is changed.