Model 6485 Picoammeter Instruction Manual DISPlay, FORMat, and SYSTem 13-7
Bit 7 (Overvoltage) Set to 1 if measurement performed with an overvoltage condition
on the input.
Bit 9 (Zero Check) Set to 1 when zero check is enabled.
Bit 10 (Zero Correct) Set to 1 when zero correct is enabled.
Example The ASCII data string contains all three data elements. The status value of
138 has a binary equivalent of 01001010, which indicates that bits B1, B3, and B7 are set.
Therefore, the reading is 1.04056uA with null (REL) and the AVG lter enabled. The
reading was taken 223.6299 seconds after the instrument was turned on.
C) FORMat:BORDer <name>
Parameters NORMal = Normal byte order for IEEE-754 binary format
SWAPped = Reverse byte order for IEEE-754 binary format
For normal byte order, the data format for each element is sent as follows:
Byte 1 Byte 2 Byte 3 Byte 4
For reverse byte order, data is sent as follows:
Byte 4 Byte 3 Byte 2 Byte 1
The #0 header is not affected by this command. The header is always sent at the begin-
ning of the data string for each measurement conversion.
The ASCII data format can only be sent in the normal byte order. The SWAPped selection
is ignored when the ASCII format is selected.
NOTE The SWAPped byte order must be used when transmitting binary data to any
IBM PC.
Bit 6 Bit 5
0 0 All limit tests passed
0 1 CALC2:LIM1 test failed
1 0 CALC2:LIM2 test failed

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)