Model 6485 Picoammeter Instruction Manual DDC Emulation Commands D-13
The various bits in the status byte are described below:
Reading Overflow Set when an overrange input is applied to the instrument.
Data Store Full Set when the number of readings stored is equal to the buffer size.
Reading Done Set when the instrument has completed the present conversion and is
ready to take another reading.
Busy The instrument is still executing a prior command and is not yet ready to accept a
new command.
IDDCO An illegal command option such as R8 has been sent. This bit is cleared when
the status byte is read.
IDDC An illegal command will set this bit. For example, S1 is illegal since no such let-
ter exists in the command set. The IDDC bit will be cleared on a reading of the status byte.
Not in remote Model 6485 in local mode of operation.
NOTE Once the Model 6485 has generated an SRQ, its status byte must be read to
clear the SRQ line. Otherwise, the instrument will continuously assert SRQ.
The Model 6485 may be programmed to generate an SRQ for more than one
condition simultaneously.
Table D-2
Status byte and mask interpretation
Bit Bit 5=0 (Data Conditions) Bit 5=1 (Error Conditions)
0 (LSB) Overow IDDCO
1 Data Store Full IDDC
2 N/A No remote
3 Reading done N/A
4 Busy N/A
5 Data Error
6 SRQ SRQ
7 N/A N/A

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)