6-6 Buffer Model 6485 Picoammeter Instruction Manual
A) TRACe:FREE?
Two values, separated by commas, are returned. The rst value indicates how many bytes
of memory are available, and the second value indicates how many bytes are reserved to
store readings.
B) TRACe:FEED <name>
Name parameters:
SENSe Raw input readings are stored in the buffer.
CALCulate1 The results of the mX+b, m/X+b, or log calculation are stored in
the buffer. See Section 5 for information on mX+b, m/X+b, or log.
CALCulate2 Test limit or Rel readings are stored in the buffer. See Section 8 for
information on limit tests.
C) TRACe:FEED:CONTrol <name>
Name parameters:
NEXT Enables the buffer and turns on the asterisk (*) annunciator. After the
buffer stores the specied number of readings, the asterisk annunciator turns off.
NEVer Disables the buffer.
D) TRACe:TSTamp:FORMat <name>
Name parameters:
ABSolute Each timestamp is referenced to the rst reading stored in the buffer.
DELTa Timestamps provide the time between each buffer reading.
The timestamp data element can be included with each buffer reading (see Ref F).
E) TRACe:DATA?
1. The response message will include one to four data elements for each stored read-
ing. Use the FORMat:ELEMents command (see Ref F) to specify the elements.
2. Reading an empty buffer will result in the ERROR: -230 display message.
3. Buffer data can be sent in the binary format. (See Table 14-3, Table , FORMat
command summary, on page 5 for details).
F) FORMat:ELEMents <list>
List parameters:
READing Includes the buffer reading in each data string.
UNITs identies amps or math units.
TIME Includes the timestamp for each reading. Timestamp can be in the abso-
lute or delta format (see Ref D).

Table of Contents

Other manuals for Keithley 6485

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 6485 and is the answer not in the manual?

Keithley 6485 Specifications

General IconGeneral
ModelKeithley 6485
CategoryPicoammeter
Input Impedance>10^14 Ohms
Voltage Burden<200µV
Accuracy0.4% of reading
Voltage Source RangeNot Applicable
Voltage Source ResolutionNot Applicable
Voltage Source AccuracyNot Applicable
InterfaceIEEE-488
Power Requirements100V to 240V AC, 50/60Hz
Dimensions89 mm high x 214 mm wide x 369 mm deep (3.5 in x 8.4 in x 14.5 in)

Related product manuals