138 Chapter 6
Replacement Procedure
Front Panel Removal
Front Panel Removal
Tools required
TORX screwdriver T10, T15 and T20
Flat edge screwdriver
Open torque wrench, 5/16 inch (set to 10 in-lb) (for reconnecting SMA connector)
Procedure
Refer to Figure 6-2 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136.
Step 2. Disconnect the two semi-rigid cables (item 1) from the front panel.
NOTE When reconnecting the semi-rigid cables, make sure that the cables are connected to the
same connectors from which they previously came.
Step 3. Remove the top trim (item 2) using a flat edge driver.
Step 4. Remove the seven TORX T15 screws (item 3 ) fastening front panel.
Step 5. Release the flat cables from the cable clamps (item 4).
Step 6. Gradually push the front panel assembly towards the outside.
Step 7. Disconnect the cable (red and black) ( item 5) from the front panel.
Step 8. Disconnect the cable (item 6) from the front panel.