34 Chapter 2
Performance Tests
E5052B Performance Test Record
@ Frequency Band: 250 MHz - 7 GHz, Power Level: +20 dBm
Power Measurement Accuracy Test
@ -20 dBm
@ -10 dBm
Frequency
[Hz]
Frequency
Resolution [Hz]
Test Limit
[Hz]
Test result
[Hz]
Measurement
uncertainty [Hz]
TAR Status
301 M 64 k ± 64 k ± 0.94 k 68.0 PASS
301 M 1 k ± 1 k ± 1.91 523.5 PASS
301 M 10 ± 10 ± 0.26 38.4 PASS
7 G 64 k ± 64 k ± 4.41 k 14.5 PASS
7 G 1 k ± 1 k ± 18.41 54.3 PASS
7 G 10 ± 10 ± 0.42 23.8 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
30 M ± 1.0 ± 0.188 5.88 PASS
1 G ± 1.0 ± 0.186 5.93 PASS
2.0125 G ± 1.0 ± 0.190 5.79 PASS
3.0125 G ± 1.0 ± 0.191 5.78 PASS
4.0125 G ± 1.0 ± 0.292 3.73 PASS
5.0125 G ± 1.0 ± 0.305 3.56 PASS
6.0125 G ± 1.0 ± 0.299 3.65 PASS
7 G ± 1.0 ± 0.299 3.65 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
30 M ± 1.0 ± 0.206 5.33 PASS
1 G ± 1.0 ± 0.161 6.88 PASS
2.0125 G ± 1.0 ± 0.183 6.03 PASS
3.0125 G ± 1.0 ± 0.196 5.62 PASS
4.0125 G ± 1.0 ± 0.288 3.78 PASS
5.0125 G ± 1.0 ± 0.286 3.81 PASS