158 Chapter 6
Replacement Procedure
Oven CH2 Assembly Replacement
Oven CH2 Assembly Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Removal Procedure
Refer to Figure 6-12 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136.
Step 2. Remove the four TORX T10 screws (item 1) fastening the oven ch2 Assembly.
Step 3. Disconnect the RF cable (item 2) from the oven ch2 Assembly.
Step 4. Disconnect the cable (item 3) from the connector.
Step 5. Release the cable (item 4) from the cable clamp.
Step 6. Disconnect the connector (item 5) from the oven ch2 Assembly.
Step 7. Lift the oven ch2 Assembly (item 6).
Step 8. Remove the three TORX T10 screws (item 7) fastening the C-bracket.
Step 9. Remove the oven ch2 (item 8).