168 Chapter 6
Replacement Procedure
A1 Freq / Power Detector Module Replacement
A1 Freq / Power Detector Module Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)
Removal Procedure
Refer to Figure 6-17 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136
Step 2. Remove the front panel as described in “Front Panel Removal” on page 138.
Step 3. Disconnect the RF cable(item 1).
Step 4. Disconnect the semi-rigid cables(item 2).
Step 5. Disconnect the semi-rigid cable(item 3).
Step 6. Disconnect the semi-rigid cable(item 4).
Step 7. Remove the TORX T15 screw fastening the freq / power detector module bracket (item 5).
Step 8. Slide out the freq / power detector module (item 6).