24 Chapter 2
Performance Tests
Individual Test Description
6. SA Relative Level Accuracy Test
Description
This test checks the relative level accuracy for the dpectrum monitor of the E5052B. The
mixed signals generated by two signal generators are used, and the carrier frequencies for
this test are 10 MHz, 1 GHz and 3 GHz.
Test Equipment
Signal Generator 1 Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
Signal Generator 2 (or Sweeper) Keysight E5070B/71B (use of the source)
Power Meter Keysight E4419B
Power Sensor Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Step Attenuator Keysight 8496H
Power Splitter Keysight 11667A
N(m)-N(m) Adapter Keysight p/n 1250-1475
Power Divider Keysight 11636A
N(m)-N(m) cable 61cm, 3 ea. Keysight N6314A (p/n 8120-8862)
BNC cable, 122 cm, 2 ea. Keysight p/n 8120-1840