180 Chapter 6
Replacement Procedure
Front Panel Cover Removal
Front Panel Cover Removal
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)
Removal Procedure
Refer to Figure 6-23 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 138.
Step 3. Disconnect the cables (item 1, 2) connected through cover.
Step 4. Remove the seven TORX T10 screws (item 3) fastening the cover.
Figure 6-23 Front Panel Cover Removal
e5052bse1091
2 1