Chapter 6 181
Replacement Procedure
Control Touch Panel Replacement
6. Replacement Procedure
Control Touch Panel Replacement
Tools Required
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)
Removal Procedure
Refer to Figure 6-24 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 138.
Step 3. Disconnect the cables (item 1, 2) connected through cover.
Step 4. Remove the two TORX T10 screws (item 3) fastening the control touch panel.
Figure 6-24 Control Touch Panel Removal
Replacement Procedure
Step 1. Replace the control touch panel by inverse procedure of removal.