Chapter 2 39
Performance Tests
E5052B Performance Test Record
2. Performance Test
CW Frequency: 3 GHz
DC Power Voltage Test
+7.5 M ± 1.5 ± 0.287 6.05 PASS
Offset Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
Offset Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
-7.5 M ± 1.5 ± 0.314 5.51 PASS
-6.5 M ± 1.5 ± 0.361 4.76 PASS
-5.5 M ± 1.5 ± 0.308 5.63 PASS
-4.5 M ± 1.5 ± 0.360 4.78 PASS
-3.5 M ± 1.5 ± 0.368 4.67 PASS
-2.5 M ± 1.5 ± 0.325 5.31 PASS
-1.5 M ± 1.5 ± 0.333 5.19 PASS
-0.5 M ± 1.5 ± 0.370 4.64 PASS
+0.5 M ± 1.5 ± 0.320 5.40 PASS
+1.5 M ± 1.5 ± 0.328 5.26 PASS
+2.5 M ± 1.5 ± 0.337 5.12 PASS
+3.5 M ± 1.5 ± 0.373 4.60 PASS
+4.5 M ± 1.5 ± 0.339 5.08 PASS
+5.5 M ± 1.5 ± 0.330 5.23 PASS
+6.5 M ± 1.5 ± 0.377 4.55 PASS
+7.5 M ± 1.5 ± 0.318 5.43 PASS
DC POWER
Voltage [V]
Test limit [mV] Test result
[mV]
Measurement
uncertainty [mV]
TAR Status
0 ± 2.0 ± 0.140 14.28 PASS
1 ± 4.0 ± 0.159 25.15 PASS
2 ± 6.0 ± 0.181 33.14 PASS
5 ± 12.0 ± 0.243 49.38 PASS
10 ± 22.0 ± 0.366 60.10 PASS
12 ± 26.0 ± 0.429 60.60 PASS