36 Chapter 2
Performance Tests
E5052B Performance Test Record
Phase Noise Accuracy Test
@ Frequency : 1 GHz
@ Frequency : 10 MHz
1 G ± 1.0 ± 0.158 3.30 PASS
2.0125 G ± 1.0 ± 0.150 3.47 PASS
3.0125 G ± 1.0 ± 0.170 6.49 PASS
4.0125 G ± 1.0 ± 0.282 3.86 PASS
5.0125 G ± 1.0 ± 0.275 3.96 PASS
6.0125 G ± 1.0 ± 0.273 3.99 PASS
7 G ± 1.0 ± 0.276 3.95 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
Offset
Frequency [Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
- 100 M ± 3.0 ± 0.231 18.22 PASS
- 40 M ± 2.0 ± 0.236 10.46 PASS
- 10 M ± 2.0 ± 0.242 10.22 PASS
- 1 M ± 2.0 ± 0.238 10.36 PASS
- 1 k ± 2.0 ± 0.237 10.42 PASS
- 100 ± 3.0 ± 0.291 14.35 PASS
100 ± 3.0 ± 0.282 14.85 PASS
1 k ± 2.0 ± 0.240 10.30 PASS
1 M ± 2.0 ± 0.246 10.06 PASS
10 M ± 2.0 ± 0.234 10.59 PASS
40 M ± 2.0 ± 0.236 10.46 PASS
100 M ± 3.0 ± 0.229 18.39 PASS
Offset
Frequency [Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
- 1 M ± 2.0 ± 0.252 9.79 PASS
- 1 k ± 2.0 ± 0.257 9.60 PASS