Chapter 2 35
Performance Tests
E5052B Performance Test Record
2. Performance Test
@ 0 dBm
@ +10 dBm
@ +20 dBm
6.0125 G ± 1.0 ± 0.287 3.79 PASS
7 G ± 1.0 ± 0.300 3.63 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
30 M ± 1.0 ± 0.196 5.62 PASS
1 G ± 1.0 ± 0.159 3.29 PASS
2.0125 G ± 1.0 ± 0.152 3.43 PASS
3.0125 G ± 1.0 ± 0.157 7.05 PASS
4.0125 G ± 1.0 ± 0.268 4.08 PASS
5.0125 G ± 1.0 ± 0.278 3.91 PASS
6.0125 G ± 1.0 ± 0.271 4.03 PASS
7 G ± 1.0 ± 0.275 3.96 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
30 M ± 1.0 ± 0.195 5.66 PASS
1 G ± 1.0 ± 0.158 3.31 PASS
2.0125 G ± 1.0 ± 0.153 3.40 PASS
3.0125 G ± 1.0 ± 0.163 6.78 PASS
4.0125 G ± 1.0 ± 0.270 4.04 PASS
5.0125 G ± 1.0 ± 0.277 3.93 PASS
6.0125 G ± 1.0 ± 0.272 4.01 PASS
7 G ± 1.0 ± 0.276 3.95 PASS
Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
30 M ± 1.0 ± 0.193 5.70 PASS