190 Chapter 6
Replacement Procedure
Probe Power and Power Switch Board Replacement
Probe Power and Power Switch Board Replacement
Tools Required
TORX screwdriver, T8, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)
Hex key, 1.5 mm
Removal Procedure
Refer to Figure 6-29 for this procedure.
Step 1. Remove the outer cover as described in “Outer Cover Removal” on page 136.
Step 2. Remove the front panel as described in “Front Panel Removal” on page 138.
Step 3. Remove the front panel cover as described in “Front Panel Cover Removal” on page 180.
Step 4. Remove the Probe Power Board or Swithc Board.
For removing the Prove Power Board.
Disconnect the cable (item 1) from the front panel keyboard.
Remove the two TORX T10 screws (item 2) fastening the front panel.
For removint the Swithc Board.
Disconnect the cable (item 3) connected the front panel keyboard.
Release the clamped cables from the clamps (item 4) on the LCD.
Cut the cable tie (item 5).
Remove the two TORX T8 screws (item 6) fastening the angle.