Chapter 3 47
Adjustment
Required Adjustment after Replacing Assembly
1. Chapter Title
2. Chapter Title 3. Adjustment
4. Chapter Title
5. Chapter Title
PM Linearity Adjustment
The purpose of this procedure is to adjust the PM linearity of the A1 freq/power detector
module.
Required equipment for the PM linearity adjustment
PM Temp Adjustment
The purpose of this procedure is to adjust the PM temp of the A1 freq/power detector
module.
Required equipment for the PM temp adjustment
None
Description Recommended Model
Signal Generator Keysight E8257D with Opt. UNX, UNT, 1EA
without Opt. UNW, 1EH
Power Amplifier (35dB) R & K A3000-2H-R
Step Attenuator Keysight 8496H
Attenuator (3dB) Keysight 8491A Opt. 003
Blocking Capacitor Keysight 11742A
Low Pass Filter Mini-Circuits SLP-70 (70MHz)
Power Splitter Keysight 11667A
Power Sensor Keysight E9304A Opt. H18 or Keysight E9304A
Opt. H19
Power Meter Keysight E4419A/B
N(m)-N(m) Cable Keysight N6314A (p/n 8120-8862)
N(m)-N(m) Adapter Keysight p/n 1250-1475
N(f)-SMA(m) Adapter Keysight p/n 1250-1562