38 Chapter 2
Performance Tests
E5052B Performance Test Record
CW Frequency: 1 GHz
-2.5 M ± 1.5 ± 0.315 5.47 PASS
-1.5 M ± 1.5 ± 0.318 5.43 PASS
-0.5 M ± 1.5 ± 0.315 5.47 PASS
+0.5 M ± 1.5 ± 0.301 5.76 PASS
+1.5 M ± 1.5 ± 0.314 5.51 PASS
+2.5 M ± 1.5 ± 0.320 5.40 PASS
+3.5 M ± 1.5 ± 0.334 5.17 PASS
+4.5 M ± 1.5 ± 0.331 5.22 PASS
+5.5 M ± 1.5 ± 0.342 5.04 PASS
+6.5 M ± 1.5 ± 0.319 5.42 PASS
+7.5 M ± 1.5 ± 0.330 5.23 PASS
Offset Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
Offset Frequency
[Hz]
Test limit [dB] Test result
[dB]
Measurement
uncertainty [dB]
TAR Status
-7.5 M ± 1.5 ± 0.302 5.73 PASS
-6.5 M ± 1.5 ± 0.327 5.28 PASS
-5.5 M ± 1.5 ± 0.304 5.69 PASS
-4.5 M ± 1.5 ± 0.308 5.63 PASS
-3.5 M ± 1.5 ± 0.325 5.31 PASS
-2.5 M ± 1.5 ± 0.318 5.43 PASS
-1.5 M ± 1.5 ± 0.316 5.48 PASS
-0.5 M ± 1.5 ± 0.301 5.76 PASS
+0.5 M ± 1.5 ± 0.314 5.51 PASS
+1.5 M ± 1.5 ± 0.320 5.40 PASS
+2.5 M ± 1.5 ± 0.334 5.17 PASS
+3.5 M ± 1.5 ± 0.331 5.22 PASS
+4.5 M ± 1.5 ± 0.342 5.04 PASS
+5.5 M ± 1.5 ± 0.319 5.42 PASS
+6.5 M ± 1.5 ± 0.330 5.23 PASS