306 M8190A User’s Guide
None
None
The command form writes one or multiple amplitude table entries starting at
<amplitude_table_index>. The query form reads <length> entries starting at
<amplitude_table_index>. The data for each entry consists of a double value for the
amplitude scale factor in the range [0.0, 1.0]. The data can be given in IEEE binary
block format or as a comma-separated list of values.
Command
:ATAB:DATA 0,1
Query
:ATAB:DATA? 0,1
8.26.2 [:SOURce]:ATABle[1|2]:RESet
:ATAB:RES
:ATABle:RESet
None
None
Reset all amplitude table entries to default values.