General Programming 8
M8190A User’s Guide 193
8.7.4.5 :STATus:OPERation:PTRansition[?]
Set or queries the positive-transition register in the operation status group. A positive
transition filter allows an event to be reported when a condition changes from false
to true. Setting both positive/negative filters true allows an event to be reported
anytime the condition changes. Clearing both filters disable event reporting. The
contents of transition filters are unchanged by *CLS and *RST.
8.7.5 Voltage Status Subsystem
The Voltage Status register contains the voltage conditions of the individual
channels.
The following SCPI commands and queries are supported:
:STATus:QUEStionable:VOLTage[:EVENt]?
:STATus:QUEStionable:VOLTage:CONDition?
:STATus:QUEStionable:VOLTage:ENABle[?]
:STATus:QUEStionable:VOLTage:NTRansition[?]
:STATus:QUEStionable:VOLTage:PTRansition[?]
Table 8-5: Voltage Status Register
Bit Number
Decimal Value
Definition
0 Voltage warning
1
Output 1 has been switched off (to protect itself)
1 Voltage warning
2
Output 2 has been switched off (to protect itself)
2 Amplitude Clipped
4
The amplitude for output 1 has been clipped to the highest/lowest possible
DAC value.
3 Amplitude Clipped
8
The amplitude for output 2 has been clipped to the highest/lowest possible
DAC value.