General Programming 8
M8190A User’s Guide 237
8.15.2 [:SOURce]:ROSCillator:SOURce:CHECk? EXTernal|AXI|INTernal
:ROSC:SOUR:CHEC?
:ROSCillator:SOURce:CHECk?
EXTernal|AXI|INTernal
None
Check if a reference clock source is available. Returns 1 if it is available and 0 if not.
Query
:ROSC:SOUR:CHEC? AXI
8.15.3 [:SOURce]:ROSCillator:FREQuency[?] <frequency>|MINimum|MAXimum
:ROSC:FREQ[?]
:ROSCillator:FREQuency[?]
<frequency>|MINimum|MAXimum
None
Set or query the expected reference clock frequency, if the external reference clock
source is selected.
Command
:ROSC:FREQ MIN
Query
:ROSC:FREQ?