8 General Programming
268 M8190A User’s Guide
8.21.10 [:SOURce]:STABle[1|2]:DYNamic:[STATe][?] OFF|ON|0|1
:STAB:DYN[?]
:STABle:DYNamic[?]
OFF | ON | 0 | 1
None
Use this command to enable or disable dynamic mode.
If dynamic mode is switched off, segments or sequences can only be switched in
program mode, that is signal generation must be stopped. In arbitrary mode use
TRACe[1|2]:SELect to switch to a new segment. In sequence mode use
[:SOURce]:STABle[1|2]:SEQuence:SELect to switch to a new sequence.
If dynamic mode is switched on, segments or sequences can be switched
dynamically when signal generation is active. The next segment or sequence is either
selected by the command [:SOURce]:STABle[1|2]:DYNamic:SELect or by a signal fed
into the dynamic port on the module front panel. The external input values select
sequence table entries with corresponding indices.
The number of valid bits of the dynamic port is specified by the
:ARM[:SEQuence][:STARt][:LAYer]: DYNPort:WIDTh[?][ LOWerbits|ALLBits] command.
Command
:STAB:DYN 0
Query
:STAB:DYN?
8.21.11 [:SOURce]:STABle[1|2]:DYNamic:SELect[?] <sequence_table_index>
:STAB:DYN:SEL[?]
:STABle:DYNamic:SELect[?]
<sequence_table_index>
None