General Programming 8
M8190A User’s Guide 191
8.7.3.5 :STATus:QUEStionable:PTRansition[?]
Set or queries the positive-transition register in the questionable status group. A
positive transition filter allows an event to be reported when a condition changes
from false to true. Setting both positive/negative filters true allows an event to be
reported anytime the condition changes. Clearing both filters disable event reporting.
The contents of transition filters are unchanged by *CLS and *RST.
8.7.4 Operation Status Subsystem
The Operation Status register contains conditions which are part of the instrument’s
normal operation.
Table 8-4: Operation Status Register
Bit Number
Decimal Value
Definition
0 Not used
1
Returns “0”
1 Not used
2
Returns “0”
2 Not used
4
Returns “0”
3 Not used
8
Returns “0”
4 Not used
16
Returns “0”
5 Not used
32
Returns “0”
6 Not used
64
Returns “0”
7 Not used
128
Returns “0”
8 Run Status
256
Indicates if system is running (:INIT:IMM was
executed and signals are generated).
9 Not used
512
Returns “0”
10 Not used
1024
Returns “0”
11 Not used
2048
Returns “0”
12 Not used
4096
Returns “0”
13 Not used
8192
Returns “0”
14 Not used
16384
Returns “0”
15 Not used
32768
Returns “0”
The following commands access the operation status group.
8.7.4.1 :STATus:OPERation[:EVENt]?
Reads the event register in the operation status group. It’s a read-only register. Once
a bit is set, it remains set until cleared by this command or *CLS command. A query
of the register returns a decimal value which corresponds to the binary-weighted
sum of all bits set in the register.