8 General Programming
200 M8190A User’s Guide
Set or query the arming mode.
Command
:INIT:CONT:ENAB SELF
Query
:INIT:CONT:ENAB?
8.8.6 :INITiate:CONTinous[1|2][:STATe][?] OFF|ON|0|1
:INIT:CONT[1|2]:STAT[?]
:INITiate:CONTinuous[1|2]:STATe[?]
OFF | ON | 0 | 1
None
Set or query the continuous mode. This command must be used together with
INIT:GATE[1|2] to set the trigger mode.
0/OFF Continuous mode is off. If gate mode is off, the trigger mode is “triggered”,
else it is “gated”.
1/ON Continuous mode is on. Trigger mode is “automatic”. The value of gate mode
is not relevant.
Command
:INIT:CONT:STAT ON
Query
:INIT:CONT:STAT?
8.8.7 :INITiate:GATE[1|2][:STATe][?] OFF|ON|0|1
:INIT:GATE[1|2]:STAT[?]
:INITiate:GATE[1|2]:STATe[?]