3 Sequencing
122 M8190A User’s Guide
3.9.1 Dynamic Continuous
The selected segment or sequence is infinitely played until a new segment/sequence is
selected which then is played instead. After a change request the actually selected
segment/sequence is played until the end (including loop counts). Then the change
towards the new segment or sequence is performed without any gap.
Limitations:
The time between two change requests of waveforms must be bigger than the
waveform length of the biggest waveform including the loop counts.
The change delay from applying changes at the dynamic port to seeing them at the
output is the trigger to output delay (see datasheet) plus 256 sync clock cycles
minimum. Due to instrument internal functionality, this delay cannot be specified
exactly and it is always possible that one more segment/sequence A is played
before switching to B.
A AAAA BBAA
Dynamic Change
At Dynamic Port
Change Delay
Figure 3-32: Dynamic Continuous