102 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–201 Invalid while in local The command cannot be executed while the instrument in Local
control.
–202 Settings lost due to rtl A “return to local” control was forced and some settings were lost
as a result of this.
–203 Command protected The command could not be executed because it is disabled. It
was disabled by licensing or password protection
–203 Command protected; feature
not licensed
The specified feature, for example “N9073A-TR2” is not licensed.
The license may have expired. You cannot use it until you get a
license.
–210 Trigger error A trigger error has occurred, but the exact problem cannot be
specifically identified.
–211 Trigger ignored A GET, *TRG or other triggering signal was received, but it was
ignored because of timing considerations. For example, maybe
the instrument was not ready to respond when the command was
received.
–212 Arm ignored An arming signal was received, but it was ignored.
–213 Init ignored An initiate trigger/sweep request was received and ignored,
because another measurement was already in progress.
–214 Trigger deadlock The trigger source for the initiation of a measurement is set to
GET, and the following measurement query was received. The
measurement cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.
–215 Arm deadlock The arm source for the initiation of a measurement is set to GET
and the following measurement query is received. The
measurement cannot be started until a GET is received and the
GET would cause an INTERRUPTED error.
–220 Parameter error A problem was found with a program data element. The exact
problem cannot be specifically identified.
–221 Settings conflict; There are many types of settings conflict errors. See section 3.5
for information about these errors.
–222 Data out of range; A data element was found but the instrument could not be set to
that value because it was outside the range defined for the
command. A descriptive message may be appended, such as
“clipped to upper limit”
–222 Data out of Range; clipped
to source max/min
A source parameter has been entered that exceeds the range of
the selected source. The parameter has been clipped to match
the range of the source
–222 Data out of range; Invalid
list data
You tried to use a trace that has a number of sweep points that is
different from the current setting of sweep points.
Err# Message Verbose/Correction Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals