EasyManuals Logo

Keysight N9040B Option 508 User Manual

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #422 background imageLoading...
Page #422 background image
422 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Assembly Replacement Procedures
Rear Panel
Replacement
1. Refer to Figure 15-71. Install the EXT REF IN connector and the AUX IF
OUT connector through the rear panel ports. Use the lock washers and
nuts to secure. Use a 9/16 inch nut driver on the EXT REF IN connector
and torque to 21 inch-pounds. Use a 5/16 inch nut driver on the AUX IF
OUT connector and torque to 10 inch-pounds. Reinstall the rear panel into
position in the chassis.
2. For Serial Number Prefix < 5616:
Refer to Figure 15-71. Replace the twenty-one screws (1)
(0515-0372) and the four screws (2) (0515-1946). Torque all
screws to 9 inch-pounds.
3. For Serial Number Prefix 5616:
Refer to Figure 15-72. Replace the fifteen screws (1) (0515-0372)
and the four screws (2) (0515-1946). Torque to 9 inch-pounds.
Replace the eight screws (3) (0515-0433). Torque to
21 inch-pounds.
4. Replace the instrument outer case. Refer to the “Instrument Outer Case”
replacement procedure.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals