100 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Instrument Messages
Event Messages
–200 to –299, Execution Errors
For –221 error messages, see the previous sections.
Note that Execution Errors are divided into subclasses:
–21x – Trigger errors
–22x – Parameter error
–23x – Data corrupt or stale (invalid data)
24x – Hardware error
–25x – Mass storage error
–26x – Expression data error
–27x – Macro error
–28x – Program error (a downloaded program-related execution error)
–29x – Memory use error
Err# Message Verbose/Correction Information
–200 All ranges are off. Turn
on at least a range
There are no range turn on in scan table. You need to turn on at
least a range to initiate a scan.
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom.
–200 At Full Zoom Marker Zoom is not available as it has reached full zoom.
–200 Execution Error A program execution error has occurred. The exact problem
cannot be specifically identified.
–200 Execution error; Carrier
frequency outside device’s
transmit band
The entered channel/carrier frequency is not within the range of
your current mode setup selection of standard and device.
–200 Execution error; Invalid
GSM burst timing
A GSM-like burst was acquired, but its timing is not valid. Ensure
the correct Burst Type has been selected.
–200 Execution error; Invalid
IP address
The IP address supplied is either not valid or does not belong to a
compatible Signal Generator. Please check the IP address and
instrument connection and try again.
–200 Execution error; Invalid
Marker Trace.
Cannot place markers on the reference trace, because the
reference trace is currently turned off or has no data.
–200 Execution Error; No peak
found.
No signal peak was found within the defined parameters of the
search.
–200 Execution error; No ranges
are defined. Activate a
range.
There are no active ranges in the range table. You will need to
activate at least one range.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight N9040B Option 508 and is the answer not in the manual?

Keysight N9040B Option 508 Specifications

General IconGeneral
BrandKeysight
ModelN9040B Option 508
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals