200                  Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Analog/Digital IF Troubleshooting
A3 Digital IF Troubleshooting
A3 Digital IF Troubleshooting
Verifying the 22.5 MHz Input Power
1. Perform an instrument shutdown. 
2. Remove the cover of the UXA Signal Analyzer. Refer to Chapter 15, 
“Assembly Replacement Procedures”, on page  347 in this service guide. 
3. Turn the instrument over so that the bottom side of the analyzer is facing 
up. 
4. Turn on the UXA Signal Analyzer and wait for the instrument to complete 
the boot up process. 
5. Press System, Alignments, Auto Align, Off. Press X
6. Press the pull down menu.
7. Press Input/Output, RF Calibrator, 50 MHz
8. Verify the 50 MHz signal is at −25 dBm by pressing FREQ, 50 MHz, SPAN, 
1 MHz, Peak Search on the analyzer. The marker readout should be 50 
MHz at −25 dBm ± 3 dB. If this reference signal is measuring incorrectly, 
see Chapter 4, “RF Section Troubleshooting (RF/Microwave Analyzers)”, 
on page  139 in this service guide.
9. To continue verifying press SPAN, Zero Span. Verify the input attenuator 
on the UXA is set to 10 dB. Look near the top of the display near the center 
and verify Atten: 10 dB is visible. If needed change the input attenuator by 
pressing AMPTD, Attenuation, 10 dB on the analyzer.