EasyManua.ls Logo

Keysight N9040B Option 508 - Before Performing a Functional Test

Keysight N9040B Option 508
542 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
504 Keysight Technologies N9040B UXA Signal Analyzer Service Guide
Functional Tests
Before Performing a Functional Test
Before Performing a Functional Test
1. Ensure that you have the proper test equipment.
2. Switch on the unit under test (UUT) and let it warm up (in accordance with
warm-up requirements in the instrument specifications).
3. Allow sufficient warm-up time for the required test equipment (refer to
individual instrument documentation for warm-up specifications).
4. Ensure that the analyzer’s frequency reference is set to Internal:
a. Press the Input/Output, Freq Ref In.
b. If the Freq Ref In softkey does not show Internal, press the Freq Ref
Input softkey and select Internal.
5. Following instrument warm-up, perform the auto align routine:
Press System, Alignments, Align Now, All.
Functional test accuracy depends on the precision of the test equipment used. Ensure that all of
the test equipment is calibrated before running a functional test.

Table of Contents

Related product manuals